Simulation and Measurement of Resistance & Pt Sensors, Current and Voltage / Li-Ion Battery / Continuity Checker / Ramp Function�
– Simulation and measurement mode
– Battery operation
– Graphic LCD
– 24V supply for current loops
– HART compatible
– Manual mode & ramp function
– Continuity test
– Data logger function
Measuring range | Resolution | Accuracy (of rdg.) |
Measuring parameter voltage DC V | ||
0 … 30V | 0.001V | ± 0.02%�of rdg. + 2 Dgt |
Measuring parameter current DC mA | ||
0 … 24-mA | 0.001-mA | ± 0.02%of rdg. + 2 Dgt |
Measurement parameter resistance | ||
0 … 400 Ω | 0.01 Ω | ± 0.02%�of rdg. + 0.01Ω |
Pt10 … Pt1000 | ||
-200 … 200°C / -328 … 392°F | Pt10 … Pt400: 0.01°C / 0.018°F | ± 0.2°C / 0.36°F |
200 … 600°C / 392 … 1112°F | Pt500 … Pt100: 0.1°C / 0.18°F | ± 0.3°C / 0.54°F |
600 … 850°C / 1112 … 1562°F | ± 0.1°C / 0.18°F | |
Ni100 | ||
-60 … 180°C / -76 … 356°F | 0.01°C / 0.018°F | ± 0.1°C / 0.18°F |
Ni120 | ||
-80 … 260°C / – 112 … 500°F | ± 0.1°C / 0.18°F | |
Cu10 | ||
-200 … 260°C / -328 … 500°F | ± 0.2°C / 0.36°F | |
*In 4-wire measuring mode, a resolution of up to 0.01 Ω in the range 0 … 1600 Ω is possible. The specified accuracy applies to the 4-wire measuring mode. With 3-wire measurement, measurement inaccuracy increases by 1°C / 1.8°F (Pt10 / Cu10), 0.6°C / 1.08°F (Pt50 / Cu50) and 0.4°C / 0.72°F (remaining types). | ||
Simulation area | Resolution | Accuracy*� |
Simulation parameter resistance | ||
0 … 400 Ω | 0.01 Ω | ± 0.02%�of rdg. + 0.01Ω |
400 … 4000 Ω | 0.1 Ω | ± 0.02%�of rdg. + 0.015Ω |
Simulation parameters Pt10 … Pt1000 | ||
-200 … 200°C / -328 … 392°F | Pt10 … Pt400: 0.01� | ± 0.15°C / 0.27°F |
200 … 600°C / 392 … 1112°F | Pt500 … Pt100: 0.1� | ± 0.25°C / 0.45°F |
600 … 850°C / 1112 … 1562°F | ± 0.15°C / 0.27°F | |
Simulation parameter Ni100 | ||
-60 … 180°C | 0.01°C / 0.018°F | ± 0.15°C / 0.27°F |
Simulation�parameter Ni120 | ||
-80 … 260°C | 0.01°C / 0.018°F | ± 0.15°C / 0.27°F |
Simulation�parameter Cu10 | ||
-200 … 260°C | 0.01°C / 0.018°F | ± 0.8°C / 1.4°F |
*Accuracy is valid at a current of> 0.2-mA or> 0.4-mA. | ||
Compatible RTD sensor | Pt10 (385), Pt50 (385), Pt100 (385), Pt200 (385), Pt400 (385), Pt500 (385), Pt1000 (385), Pt10 (3926) | |
Ni100 (672), Ni (618), Ni120 (672), Cu10 (427), Cu50 (427), Cu100 (427) | ||
General Specifications PCE-RTD 20 | ||
Display modes | Measurement: mA / V / Ω / RTD Simulation: Ω / RTD |
|
Temperature units | ° C / ° F / K | |
Current of RTD measurement | About 300 μA | |
Maximum current simulation | 3-mA (0 … 650 Ω) I < 2V / Rsim (650 … 4000 Ω) |
|
Maximum input voltage | 30V DC | |
Temperature coefficient | <30 ppm | |
Input impedance | Voltage measurement:> 1 MΩ Current measurement: 10 Ω |
|
Response time | <100 ms | |
Refresh rate display | 10 Hz | |
Data storage | Internal memory� 150000 readings |
|
Interface | USB 2.0 | |
Display | 2.4″ TFT LCD 240 x 320 pixels LED illuminated |
|
Output voltage current loop | 24V DC / 24-mA | |
HART mA loop resistance | 250 Ω ± 20% | |
Special features | Step and ramp function Automatic and manual mode √x, x2: For the measuring function |
|
Continuity test | Adjustable threshold up to 100 Ω | |
Power supply | 3.7V / 2300-mAh Li-ion battery | |
Charging time | About 5 h | |
Power adapter | Input: 100 … 240V AC / 50/60 Hz Output: 5V / 1 A DC |
|
Battery life | Approx. 15 h: Simulation and measurement with low LCD illumination Approx. 8 h: Measurement with low LCD illumination |
|
Dimensions | 162 x 82 x 40 mm / 6.4 x 3.2 x 1.6 in | |
Weight | About 300 g / < 1 lb | |
Degree of protection | IP20 | |
Operating conditions | Battery operation: 0 … 55°C / 32 … 131°F, 30 … 90% RH Mains operation: 0 … 45°C / 32 … 113°F, 30 … 90% RH | |
Storage conditions | -20 .. 60°C / -4 … 140°F, 30 … 90% rh non-condensing | |
Heating time | About 15 minutes |