Voltmeter with transistor test / Square function generation / Large LCD display / Voltage test up to 1000V / Protective rubber cover / Diode test / Continuity test / Current measurement up to 20 A.
– TRMS measurements AC / DC
– NPN and PNP transistor testing
– Rubberized housing with tripod holder
– Generation of a quadratic function
– Automatic shutdown
– Continuity test and diode test
DC | |||
Measuring range | Resolution | Accuracy | |
200 mV | 100 µV | ± (0.5% +5 digits) | |
2V | 1 mV | ± (0.5% +3 digits) | |
20V | 10 mV | ± (0.5% +3 digits) | |
200V | 100 mV | ± (0.5% +3 digits) | |
1000V | 1V | ± (1.0% +10 digits) | |
Input impedance: 10 MΩ | |||
Surge protection: 1000V DC / AC RMS | |||
AC | |||
Measuring range | Resolution | Accuracy | |
2V | 1 mV | ± (0.8% +5 digits) | |
20V | 10 mV | ± (0.8% +5 digits) | |
200V | 100 mV | ± (0.8% +5 digits) | |
1000V | 1V | ± (1.2% +10 digits) | |
Input impedance: 10 MΩ | |||
Surge protection: 1000V DC / AC RMS | |||
The accuracies refer to 10 … 100% of the respective measuring ranges, one | |||
Frequency of 40 … 400 Hz with a sine wave (True RMS) and a crest factor | |||
(Crest factor): CF < 3. With a crest factor CF > 2 the increases | |||
1% accuracy. | |||
Capacity | |||
Measuring range | Resolution | Accuracy | |
6 nF | 1 pF | ± (5.0% +40 digits) | |
60 nF | 10 pF | ± (2.5% +20 digits) | |
600 nF | 100 pF | ± (2.5% +20 digits) | |
6 µF | 1 nF | ± (2.5% +20 digits) | |
60 µF | 10 nF | ± (2.5% +20 digits) | |
600μF | 100 nF | ± (5.0% +10 digits) | |
6 mF | 1 µF | ± (5.0% +40 digits) | |
20 mF | 10 µF | ± (5.0% +40 digits) | |
Surge protection: 600V DC / AC RMS | |||
The accuracies refer to 10 … 100% of the respective measuring ranges | |||
Response time for capacities> 1 mF approx. 8 seconds | |||
The accuracy does not include the loading capacity | |||
Transistor test | |||
Measuring range | Resolution | Accuracy | |
0 … 1000 hFE | 1 hFE | – | |
For NPN and PNP transistors | |||
Base current approx. 10 µA | |||
Collector-emitter voltage approx. 3V | |||
Direct current | |||
Measuring range | Resolution | Accuracy | Voltage recording |
20 µA | 0.01 µA | ± (1.2% +8 digits) | 1.25 mV / µA |
200 µA | 0.1 µA | ± (1.2% +8 digits) | 0.125 mV / µA |
2 mA | 1 µA | ± (1.2% +8 digits) | 125 mV / mA |
20 mA | 10 µA | ± (1.2% +8 digits) | 3.75 mV / mA |
200 mA | 100 µA | ± (1.2% +8 digits) | 3.75 mV / mA |
2 A | 1 mA | ± (1.5% +10 digits) | 37.5 mV / A |
20 A. | 10 mA | ± (2.0% +6 digits) | 37.5 mV / A |
Fuse up to the measuring range 200-mA: 200-mA / 250V | |||
Fuses from a measuring range of 2 A: 20 A / 250V | |||
Alternating current | |||
Measuring range | Resolution | Accuracy | Voltage recording |
200-mA | 100 µA | ± (1.5% +15 digits) | 3.75 mV / mA |
2 A | 1-mA | ± (2.0% +5 digits) | 37.5 mV / A |
20 A. | 10-mA | ± (3.0% +10 digits) | 37.5 mV / A |
Fuse up to the measuring range 200-mA: 200-mA / 250V | |||
Fuses from a measuring range of 2 A: 20 A / 250V | |||
The accuracies refer to 10 … 100% of the respective measuring ranges, | |||
a frequency of 40 … 400 Hz with a sine wave (True RMS) and one | |||
Crest factor: CF < 3. With a crest factor CF > 2 the increases | |||
1% accuracy. | |||
Resistance | |||
Measuring range | Resolution | Accuracy | Short circuit current |
200 Ω | 0.1 Ω | ± (0.8% +5 digits) | approx.0.4-mA |
2 kΩ | 1 Ω | ± (0.8% +3 digits) | approx. 100 µA |
20 kΩ | 10 Ω | ± (0.8% +3 digits) | approx. 10 µA |
200 kΩ | 100 Ω | ± (0.8% +3 digits) | approx. 1 µA |
2 MΩ | 1 kΩ | ± (0.8% +3 digits) | approx. 0.2 µA |
200 MΩ | 100 kΩ | ± (5% +30 digits) | approx. 0.2 µA |
Voltage with open circuit: approx. 1V | |||
Surge protection: 600V DC / AC RMS | |||
Diode test | Test current | Test voltage | Accuracy |
Measuring range | Approx.0.4-mA � | Approx.3.3V | 5% |
Surge protection: 600 V DC / AC RMS | |||
Continuity test | Test current | Test voltage | Accuracy |
Measuring range | Approx.0.4-mA | Approx.3.3V | 5% |
Surge protection: 600 V DC / AC RMS | |||
Beep at 50 ± 20Ω | |||
Function generation | Tension | Function generation | |
(square) | Approx.3.3V | 50 Hz, 100 Hz, 200 Hz, 500 | |
Hz, 1 kHz, 2 kHz | |||
5 kHz, 10 kHz | |||
Surge protection: 600V DC / AC RMS | |||
Further specification | |||
Display | LCD display | ||
Largest display | 1999 (3 1/2) | ||
Measuring rate | Approx. 3 measurements per second | ||
Overrange � | “OL” display | ||
Operating conditions� | 0 … 40°C / 32 … 104°F, < 75% RH non-condensing | ||
Power supply | 1 x 9V block battery | ||
Dimensions | 190 x 88.5 x 27.5 mm | ||
Weight | Approx. 320 g / < 1 lb (with battery) | ||