The conductivity meter PCE-PHD-1-KIT1 is a true all-rounder for checking water quality. The portable conductivity meter is used to control the water parameters pH, redox, conductivity, salinity, oxygen and is also very suitable for temperature measurement.
A 2-point calibration and an automatic temperature compensation guarantees high accuracy even with highly fluctuating measuring temperatures. The data logging meter is supplied with pH and conductivity probe. Different probes are available as accessories. All values can be stored directly on the SD card (1 … 16 GB) as an Excel file or transferred via the RS-232 interface directly from the device to a PC. For this we optionally offer software and the suitable RS-232 interface cable as an accessory.
– optimal price/performance ratio
– real-time datalogger
– SD memory card (1 … 16 GB)
– the stored data is saved directly as an Excel file on the SD card (no software needed for evaluation)
– RS-232 interface for online data transmission
– measures pH, redox (optional), conductivity, salinity, oxygen, and temperature
– high precision
– low-battery display
– large LCD display
– adjustable measuring rate
– easy to use
– robust housing
– incl. pH and conductivity electrode
– automatic calibration
– manual or autom. temperature compensation
– BNC connector
– suitable for laboratory and on-site measurement
– REDOX Elektrode (optional)
pH Measurement | automatic 10 … 40 °C / 50 … 104 °F |
Measurement range | 0.00 … 14.00 pH |
Resolution | 0.01 pH |
Accuracy | ± 0.02 pH +2 digits |
Calibration | 2-point calibration (pH4, pH7) |
Temperature compensation | Automatic with additional temperature sensor (0 … 60 ºC / 32 … 140 ºF) or manual (0 … 100 ºC / 32 … 212 ºF) |
RedoxMeasurement | |
Measurement range | 1999 … 0 … 1999 mV |
Resolution | 1 mV |
Accuracy | ± 0.5 % + 2 digits |
ConductivityMeasurement | |
Measurement range | 0 … 200 μS/cm 0.2 … 2.0 mS/cm 2 … 20.0 mS/cm 20 … 200 mS/cm |
Resolution | 0.1 μS/cm 0.001 mS/cm 0.01 mS/cm 0.1 mS/cm |
Accuracy | ± 2 % of the measurement range + 1 digit |
Calibration | at 1413 μS/cm |
Temperature compensation | Automatic (0 … 60 ºC / 32 … 140 ºF) |
Total dissolved solids (TDS) Measurement | |
Measurement range | 0 … 132 ppm 132 … 1320 ppm 1320 … 13200 ppm 13200 … 132000 ppm |
Resolution | 0.1 ppm 1 ppm 10 ppm 100 ppm |
Accuracy | ± 2 % of the measurement range + 1 digit |
Temperature compensation | Automatic (0 … 60 ºC / 32 … 140 ºF) |
Salt Content Measurement | |
Measurement range | 0 … 20 mg/l (in water) 0 … 100 % (in air) 0 … 50 °C / 32 … 122 ºF |
Resolution | 0.1 mg/l 0.1 % 0.1 °C |
Accuracy | ± 0.4 mg/l ± 0.7 % ± 0.8 °C |
Calibration | In the air |
Temperature compensation | Automatic (0 … 50 ºC / 32 … 122 ºF) |
Electrode usable in the pH range | pH 4 … pH 10 |
Temperature Measurement | |
Measurement range | 0 … 60 °C/ 32 … 140 ºF |
Resolution | 0.1 °C |
Accuracy | ± 0.8 °C |
General Specifications | |
Measurement rate | adjustable from 1 second to 8 hours 59 minutes 59 seconds |
Display | LC-Display 52 x 38 mm / 2 x 1.5″ |
Memory | flexible via SD card memory 1 … 16 GB |
Interface | RS-232 |
Software | available as accessory (optional) |
Power supply | 6 x 1.5 V AA batteries / 9 V-power adapter (optional) |
Operating conditions | 0 … +50 °C / < 85 % r.H. |
Dimensions | 177 x 68 x 45 mm / 7.0 in x 2.7 in x 1.8 in |
Weight | 490 g / 1.1 lbs |